FDA recall Z-0697-2021

Haag-Streit USA Inc · Class II · device

Product

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Reason for recall

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye

Distribution

US Nationwide distribution in the state of OK.

Key facts

Status
Ongoing
Initiation date
2020-11-20
Report date
2020-12-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mason, OH, United States