Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 7627 SCLP, 7628 SCM, 7629 SCF, 7630 SCX: 7607 SSM/UDI: Natrelle Inspira SSM 140cc BI (US) 10888628032835, Natrelle Inspira SSM 175cc BI (US) 10888628032842, Natrelle Inspira SSM 195cc BI (US) 10888628032859, Natrelle Inspira SSM 210cc BI (US) 10888628032866, Natrelle Inspira SSM 240cc BI (US) 10888628032873, Natrelle Inspira SSM 255cc BI (US) 10888628032880, Natrelle Inspira SSM 275cc BI (US) 10888628032897, Natrelle Inspira SSM 295cc BI (US) 10888628032903, Natrelle Inspira SSM 310cc BI (US) 10888628032910, Natrelle Inspira SSM 330cc BI (US) 10888628032927, Natrelle Inspira SSM 345cc BI (US) 10888628032934, Natrelle Inspira SSM 360cc BI (US) 10888628032941, Natrelle Inspira SSM 375cc BI (US) 10888628032958, Natrelle Inspira SSM 405cc BI (US) 10888628032965, Natrelle Inspira SSM 445cc BI (US) 1
The firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
US:AL, AR,AZ,CA,CO,DC,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WY
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0701-2020