FDA recall Z-0716-2020

Arrow International Inc · Class II · device

Product

Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J

Reason for recall

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Distribution

U.S. Nationwide, Canada

Key facts

Status
Terminated
Initiation date
2019-11-15
Report date
2019-12-25
Termination date
2020-06-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Reading, PA, United States

Primary source