# FDA recall Z-0719-2026

> **PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.** · Class II · device recall initiated 2025-10-02.

## Product

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system.  Model Number: 722234;

## Reason for recall

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.

## Distribution

Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;

## Key facts

- **Recall number:** Z-0719-2026
- **Recalling firm:** PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-10-02
- **Report date:** 2025-12-03

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Best, N/A, Netherlands

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0719-2026

## Citation

> AI Analytics. FDA recall Z-0719-2026. Retrieved 2026-07-20 from https://api.ai-analytics.org/recall/Z-0719-2026. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*
