FDA recall Z-0722-2024

Cook Medical Llc · Class II · device

Product

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Reason for recall

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Distribution

Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM

Key facts

Status
Ongoing
Initiation date
2023-12-07
Report date
2024-01-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States