FDA recall Z-0724-2023

DeVilbiss Healthcare LLC · Class II · device

Product

DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q

Reason for recall

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Distribution

US Nationwide. Foreign distribution pending.

Key facts

Status
Ongoing
Initiation date
2022-11-18
Report date
2022-12-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Somerset, PA, United States

Primary source