FDA recall Z-0725-2026

Enterix, Inc. · Class II · device

Product

InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.

Reason for recall

InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.

Distribution

Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.

Key facts

Status
Ongoing
Initiation date
2025-09-17
Report date
2025-12-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Edison, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0725-2026