# FDA recall Z-0727-2022

> **B. Braun Medical, Inc.** · Class II · device recall initiated 2022-01-31.

## Product

Dispensing Pins for Air Inlet Filter, Product Catalog No. 412005, DP2500  Dispensing Pin W/Valve.  A valve for aspiration or injection of medication or enteral feeding solution from inverted multi-dose vials.

## Reason for recall

This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the dose of medication delivered to the patient, cause delays in treatment due to the need to replace the device or may cause contamination of the fluid path that can lead to bloodstream infection.

## Distribution

US Nationwide distribution in the states of AZ, CA, FL, LA, MI, MO, MS, NC, NJ, OH, PA, TN, TX, and UT. The country of Canada.

## Key facts

- **Recall number:** Z-0727-2022
- **Recalling firm:** B. Braun Medical, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2022-01-31
- **Report date:** 2022-03-16
- **Termination date:** 2024-04-29

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Allentown, PA, United States