FDA recall Z-0729-2022

Maquet Cardiovascular, LLC · Class II · device

Product

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. Product Code/REF Number: M002020952070

Reason for recall

Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm

Distribution

US Nationwide distribution in the state of MO.

Key facts

Status
Terminated
Initiation date
2022-01-14
Report date
2022-03-16
Termination date
2023-09-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0729-2022