FDA recall Z-0735-2021

Boston Scientific Corporation · Class I · device

Product

Lotus Edge Valve System, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System, 25mm, sterile, REF H749LVS250, GTIN 08714729940821, Made in Ireland, Ballybrit Business Park, Galway, Ireland.

Reason for recall

Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.

Distribution

Distribution was made to AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. Government distribution was also made. Foreign distribution was made to Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, South Korea, Spain, Sweden, and Switzerland.

Key facts

Status
Terminated
Initiation date
2020-11-17
Report date
2021-01-20
Termination date
2022-01-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Maple Grove, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0735-2021