FDA recall Z-0742-2025

Noah Medical · Class II · device

Product

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Reason for recall

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

Distribution

US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

Key facts

Status
Ongoing
Initiation date
2024-10-28
Report date
2025-01-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Carlos, CA, United States