FDA recall Z-0752-2023

GE Medical Systems, LLC · Class II · device

Product

Vivid S5 Ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).

Reason for recall

Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, AZERBAIJAN, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, CHINA Colombia, CONGO, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, CZECH REPUBLIC, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, HONG KONG, Hungary, ICELAND, INDIA, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Kuwait, KYRGYZSTAN, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, LUXEMBOURG, Macedo

Key facts

Status
Ongoing
Initiation date
2022-11-10
Report date
2022-12-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0752-2023