FDA recall Z-0755-2021

Roche Diagnostics Operations, Inc. · Class II · device

Product

Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001

Reason for recall

Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2020-12-08
Report date
2021-01-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States