Juno DRF, Model Number 709020 - Product Usage: General. multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.