FDA recall Z-0765-2025

MEDLINE INDUSTRIES, LP - Northfield · Class I · device

Product

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NE

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Distribution

US Nationwide Distribution

Key facts

Status
Ongoing
Initiation date
2024-12-16
Report date
2025-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States