FDA recall Z-0773-2019

NEUSOFT MEDICAL SYSTEMS IMP & EX · Class II · device

Product

NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for recall

For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.

Distribution

US Distribution in states of: KS, OK and MA.

Key facts

Status
Terminated
Initiation date
2019-01-09
Report date
2019-02-13
Termination date
2022-08-25
Voluntary/Mandated
FDA Mandated
Location
Shenyang, China

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0773-2019