FDA recall Z-0776-2018

Philips Electronics North America Corporation · Class II · device

Product

Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Reason for recall

Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.

Distribution

Domestic and Foreign

Key facts

Status
Terminated
Initiation date
2018-01-03
Report date
2018-03-07
Termination date
2019-11-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0776-2018