FDA recall Z-0776-2023

Medtronic MiniMed · Class II · device

Product

Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system

Reason for recall

An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not receive SMS notifications (could result in hypoglycemia or hyperglycemia), and sensor glucose values will also not be sent to the InPen app.

Distribution

OUS: Austria, Belgium, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Luxembourg, Moldova, Republic of, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom

Key facts

Status
Ongoing
Initiation date
2022-11-21
Report date
2023-01-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0776-2023