Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application
During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.