FDA recall Z-0782-2020

Siemens Medical Solutions USA, Inc · Class II · device

Product

Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application

Reason for recall

During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.

Distribution

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.

Key facts

Status
Terminated
Initiation date
2019-11-25
Report date
2020-01-15
Termination date
2020-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States