FDA recall Z-0801-2019

Radiometer America Inc · Class II · device

Product

Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and ABL837. ABL800 Series Ref. 393-803, 393-804, 393-805, 393-806, 393-820, 393-821, 393-825, 393-826, 393-835, 393-836, 393-844, 393-845.

Reason for recall

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.

Distribution

US: AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, PA, TN, TX, VA, WA, WI, and WV OUS: Canada, Croatia, Finland and Russia

Key facts

Status
Terminated
Initiation date
2018-07-02
Report date
2019-02-13
Termination date
2024-04-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States