FDA recall Z-0801-2021

Ortho-Clinical Diagnostics, Inc. · Class II · device

Product

VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.

Reason for recall

A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.

Distribution

Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2020-11-20
Report date
2021-01-13
Termination date
2023-10-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States

Primary source