FDA recall Z-0803-2021

Roche Diagnostics Operations, Inc. · Class II · device

Product

CoaguChek XS Prof (Professional) Meters- IVD professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. Catalog: 04837975001 Units could consist of any of the following strip configurations: 6 count, 24 count, 2x 24 count (i.e. 48 strips)* *CoaguChek XS 3x48 promo LTC HH Catalog: 06298176001 CoaguChek XS 6x24 promo LTC HH Catalog:08468699001 CoaguChek XS 4x48 promo Catalog: 08468745001 CoaguChek XS 6x48 promo Catalog: 08468753001

Reason for recall

Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results

Distribution

US Nationwide

Key facts

Status
Terminated
Initiation date
2020-12-02
Report date
2021-01-13
Termination date
2022-06-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0803-2021