FDA recall Z-0803-2025

Beckman Coulter, Inc. · Class II · device

Product

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reason for recall

Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2024-11-26
Report date
2025-01-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States