FDA recall Z-0804-2020

Flexicare Medical Ltd. · Class II · device

Product

BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320

Reason for recall

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Distribution

US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK

Key facts

Status
Terminated
Initiation date
2019-11-27
Report date
2020-01-22
Termination date
2021-04-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rhondda Cynon Taff, United Kingdom

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0804-2020