FDA recall Z-0806-2025

Elucent Medical Inc · Class II · device

Product

EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion is a component of the EnVisio Navigation System which is a medical device that detects the presence of the SmartClip Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery.

Reason for recall

There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.

Distribution

US Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA, SC, WI

Key facts

Status
Ongoing
Initiation date
2024-11-18
Report date
2025-01-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Eden Prairie, MN, United States

Primary source