FDA recall Z-0809-2022

Biomet, Inc. · Class II · device

Product

BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Reason for recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

Distribution

Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands

Key facts

Status
Terminated
Initiation date
2022-02-14
Report date
2022-03-30
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0809-2022