FDA recall Z-0812-2022

Biomet, Inc. · Class II · device

Product

BIOMET TRAUMA Drill Bit, Tap, Guide Pin: [DVR Drill Bit, Item # 212000022, DVR 2.2MM DRILL BIT], [ALPS Drill Bit, Item # 214227160, 2.7X160MM CALIBRATED DRILL], [ALPS Drill Bit, Item # 824273100, 4.5X150MM DRILL], [ALPS Drill Bit, Item # 829029170, 2.5X140MM DRILL], [ALPS Drill Bit, Item # 829032170, 3.5X140MM DRILL], [ALPS Tap, Item # 824150170, 3.5X140MM CORTICAL TAP], [ALPS Guide Pin, Item # 829516150, 1.6X150MM BAYONET TIP GUIDE PIN], [ALPS Guide Pin, Item # 829516151, 1.6X150MM THREADED TIP GUIDE PIN]. Instruments are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Reason for recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

Distribution

Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands

Key facts

Status
Terminated
Initiation date
2022-02-14
Report date
2022-03-30
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States