FDA recall Z-0816-2020

Physio-Control, Inc. · Class II · device

Product

LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected keypad.

Reason for recall

Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button.

Distribution

Worldwide Distribution - U.S Nationwide: GA, PA, IN, CA, KY, FL, AZ, WA, NC, MD, TX, CO, VT, MI, MA, TN, OH, CT, NY, MO, VA, IA, MN, LA, WV, MT, IL, AR, UT, NJ, OK, NV, NH, MS, NM, WI, OR, ID, SC, NE, AK, SD, AL, WY, RI, KS, ME, DE, ND, DC, HI, PR O.U.S. (Foreign): Germany, Netherlands, Hong Kong, Costa Rica, Canada, Singapore, Australia, France, Croatia, Italy , Sweden, Hungary,United Arab Emirates, Qatar, Ireland, Greece, Spain, Poland, Saudi Arabia, Reunion, Israel, United Kingdom, Mexico, South Africa, China, Switzerland, Denmark, Cayman Islands, Finland, Norway, Austria, Japan, Belgium, Czech Republic, New Zealand, Slovenia, Republic of Korea, Oman, Egypt, French Guiana, Iceland, Portugal, Latvia, Lebanon, Chile, Kuwait, Bosnia and Herzegovina, Yemen, Eritrea, Martinique, New Caledonia, Kenya, Romania, Guinea, Luxembourg, Kazakhstan, Papua New Guinea, Brazil, Venezuela, Bolivarian Republic of, Trinidad and Tobago, Algeria, Turkey, Ukraine, Bahrain, Congo, the Democratic Republi

Key facts

Status
Terminated
Initiation date
2019-12-20
Report date
2020-01-22
Termination date
2022-10-04
Voluntary/Mandated
Voluntary: Firm initiated