FDA recall Z-0824-2019

Steris Corporation · Class II · device

Product

ATLAS Loading Car

Reason for recall

The ATLAS loading car may not remain fully engaged with the latch on the transfer carriage when the loading car is empty, potentially resulting in the loading car unexpectedly sliding forward or falling to one side.

Distribution

Nationwide domestic distribution. Foreign distribution to Canada, Costa Rica, Indonesia, South Korea, and Taiwan.

Key facts

Status
Terminated
Initiation date
2018-12-26
Report date
2019-02-20
Termination date
2020-06-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mentor, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0824-2019