FDA recall Z-0830-2025

Abiomed, Inc. · Class I · device

Product

Impella RP Flex with SmartAssist; Product Number: 1000323;

Reason for recall

Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.

Distribution

Domestic: Nationwide Distribution International: Canada

Key facts

Status
Ongoing
Initiation date
2024-12-12
Report date
2025-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Danvers, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0830-2025