FDA recall Z-0840-2018

Philips Medical Systems (Cleveland) Inc · Class II · device

Product

Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Reason for recall

Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card; Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed; Software errors that may result in partial set of images or inability to generate CT images.

Distribution

Worldwide Distribution - US Nationwide

Key facts

Status
Terminated
Initiation date
2018-01-03
Report date
2018-03-14
Termination date
2019-07-02
Voluntary/Mandated
Voluntary: Firm initiated