FDA recall Z-0844-2020

Arrow International Inc · Class II · device

Product

Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04020-PMC The Arrow¿ Arterial Catheterization device permits access to the peripheral arterial circulation or to other small vessels.

Reason for recall

The lidstock label incorrectly displays a MR (Magnetic Resonance ) Conditional symbol (instead of a MR Unsafe symbol) in reference to a component of the kit

Distribution

US distribution to SC

Key facts

Status
Ongoing
Initiation date
2019-12-12
Report date
2020-01-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Reading, PA, United States