FDA recall Z-0846-2024

XTANT Medical Holdings, Inc · Class II · device

Product

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Reason for recall

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Distribution

US distribution in MI.

Key facts

Status
Ongoing
Initiation date
2023-11-28
Report date
2024-02-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Belgrade, MT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0846-2024