# FDA recall Z-0850-2018

> **Philips Electronics North  America Corporation** · Class II · device recall initiated 2017-10-05.

## Product

Philips Network Firewall (Cisco ASA 5506), Model # 867098

## Reason for recall

Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX.  The defect involves the clock signal component within the firewall. This component has a high probability  of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable.  This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.

## Distribution

US Nationwide

## Key facts

- **Recall number:** Z-0850-2018
- **Recalling firm:** Philips Electronics North  America Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-10-05
- **Report date:** 2018-03-14
- **Termination date:** 2019-07-01

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Andover, MA, United States