FDA recall Z-0854-2022

Illumina, Inc. · Class II · device

Product

Illumina NextSeq 550Dx, REF: 20005715, CE, IVD

Reason for recall

There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

Distribution

US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federation Saudi Arabia Singapore Slovakia South Africa South Korea Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom Vietnam

Key facts

Status
Completed
Initiation date
2022-02-22
Report date
2022-04-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0854-2022