# FDA recall Z-0863-2024

> **Smiths Medical ASD Inc.** · Class I · device recall initiated 2023-12-19.

## Product

smiths medical Medfusion Model 3500 Syringe pump

## Reason for recall

Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values.    Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.

## Distribution

worldwide

## Key facts

- **Recall number:** Z-0863-2024
- **Recalling firm:** Smiths Medical ASD Inc.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-12-19
- **Report date:** 2024-02-07

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Minneapolis, MN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0863-2024