FDA recall Z-0874-2022

Philips North America Llc · Class II · device

Product

Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0

Reason for recall

Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

Distribution

Global Distribution. US Nationwide.

Key facts

Status
Ongoing
Initiation date
2022-02-02
Report date
2022-04-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0874-2022