FDA recall Z-0875-2020

GE Healthcare, LLC · Class II · device

Product

SIGNA Architect Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

Reason for recall

It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.

Distribution

Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2019-12-23
Report date
2020-02-05
Termination date
2020-12-11