# FDA recall Z-0877-2020

> **GE Healthcare, LLC** · Class II · device recall initiated 2019-12-23.

## Product

SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

## Reason for recall

It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.

## Distribution

Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

## Key facts

- **Recall number:** Z-0877-2020
- **Recalling firm:** GE Healthcare, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-12-23
- **Report date:** 2020-02-05
- **Termination date:** 2020-12-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Waukesha, WI, United States