FDA recall Z-0877-2021

Roche Molecular Systems, Inc. · Class II · device

Product

uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001

Reason for recall

When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.

Distribution

Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.

Key facts

Status
Terminated
Initiation date
2020-11-23
Report date
2021-01-27
Termination date
2023-05-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Branchburg, NJ, United States