FDA recall Z-0880-2021

Abbott Point Of Care Inc. · Class II · device

Product

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Reason for recall

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Distribution

Nationwide

Key facts

Status
Terminated
Initiation date
2020-12-22
Report date
2021-01-27
Termination date
2023-09-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Princeton, NJ, United States