Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655TPV, AU1054096, AU1054096B, 1054096B, 1054097, BR1054096, LA1054096, CN1054096, KO1054096, KR1054096, R1054655TP, R1054655B, 1054655, 1054655TPV, 1054655TP, 1054655B
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
US Nationwide and Global Distribution to countries of: Australia, Brazil, Canada, China, Germany, Japan, Korea, and Latin America.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0882-2023