# FDA recall Z-0887-2026

> **Bard Peripheral Vascular Inc** · Class II · device recall initiated 2025-11-06.

## Product

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

## Reason for recall

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

## Distribution

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

## Key facts

- **Recall number:** Z-0887-2026
- **Recalling firm:** Bard Peripheral Vascular Inc
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-11-06
- **Report date:** 2025-12-17

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Tempe, AZ, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0887-2026