FDA recall Z-0899-2025

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · Class II · device

Product

GE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3) H2799YC; 4) H2799YD; 5) H3000WC; 6) H3000WD; 7) H3000WT; 8) H3000WW; 9) H3000WY; 10) H3000WZ; System, Tomography, Computed, Emission

Reason for recall

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2024-12-18
Report date
2025-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tirat Carmel, Israel

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0899-2025