FDA recall Z-0905-2025

Olympus Corporation of the Americas · Class I · device

Product

Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.

Reason for recall

Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.

Distribution

US Nationwide Distribution

Key facts

Status
Ongoing
Initiation date
2024-12-18
Report date
2025-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Center Valley, PA, United States