FDA recall Z-0906-2023

DeRoyal Industries Inc · Class II · device

Product

DeRoyal Tonsillectomy Tray, REF 89-10698.01

Reason for recall

DeRoyal is recalling certain lots of surgical procedure pack products that contain a ConMed Foot Controlled Electrosurgical Suction Coagulator. The recall is due to ConMed receiving reports that the suction ports may be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use.

Distribution

Distribution in US states of VA, and SC.

Key facts

Status
Terminated
Initiation date
2022-11-28
Report date
2023-01-11
Termination date
2026-03-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Powell, TN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0906-2023