FDA recall Z-0913-2020

Beckman Coulter Inc. · Class II · device

Product

iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Product Usage: used to automate the urine chemistry analysis profile using iChem¿ VELOCITY Urine Chemistry Strips.

Reason for recall

Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).

Distribution

Worldwide distributions.

Key facts

Status
Terminated
Initiation date
2019-10-28
Report date
2020-02-05
Termination date
2026-04-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0913-2020