# FDA recall Z-0915-2026

> **GE Medical Systems China Co., Ltd.** · Class I · device recall initiated 2025-11-14.

## Product

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

## Reason for recall

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure.    Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over  to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation,   manual ventilation, and volatile agent delivery may occur.   Following the reboot, the system will not return to the previous ventilation settings.

## Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.

## Key facts

- **Recall number:** Z-0915-2026
- **Recalling firm:** GE Medical Systems China Co., Ltd.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-11-14
- **Report date:** 2025-12-24

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Wuxi, China

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0915-2026

## Citation

> AI Analytics. FDA recall Z-0915-2026. Retrieved 2026-09-04 from https://api.ai-analytics.org/recall/Z-0915-2026. Source: US FDA. Licensed CC0.

---