FDA recall Z-0917-2020

Bard Peripheral Vascular Inc · Class II · device

Product

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444

Reason for recall

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.

Distribution

US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None

Key facts

Status
Terminated
Initiation date
2019-08-14
Report date
2020-02-05
Termination date
2021-02-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source