# FDA recall Z-0920-2026

> **Mindray DS USA, Inc. dba Mindray North America** · Class II · device recall initiated 2025-11-18.

## Product

Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1    HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)

## Reason for recall

When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.

## Distribution

Worldwide distribution - US Nationwide and the country of Canada.

## Key facts

- **Recall number:** Z-0920-2026
- **Recalling firm:** Mindray DS USA, Inc. dba Mindray North America
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-11-18
- **Report date:** 2025-12-24

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Mahwah, NJ, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0920-2026

## Citation

> AI Analytics. FDA recall Z-0920-2026. Retrieved 2026-09-04 from https://api.ai-analytics.org/recall/Z-0920-2026. Source: US FDA. Licensed CC0.

---